A complete, opinionated guide to authoring OpenHTF test scripts that run well in the Guppi environment — the contract, the injected vocabulary, the driver API a test author actually touches, safety layering, performance rules, and a deep section on driving the Keysight MP4300 Solar Array Simulator as an irradiance source.

This is the practitioner's companion to Instrument drivers, which covers the driver/instrument layer.


0. Mental model — what actually runs#

A Guppi test is a plain Python file in the rig's workspace under tests/. There is no test registry and no framework subclass to inherit. The rack:

  1. Reads the file from its own disk (the workspace) — the script is never shipped over the wire with the run command; test_path names a file and the rack reads it, so an edit between plan-approval and start fails loudly instead of running different code silently.
  2. exec()s it in a locked-down namespace (see §3).
  3. Pulls the module-level TEST_PHASES list out of the namespace.
  4. Builds htf.Test(*TEST_PHASES) and calls test.execute(...).
  5. Binds every configured rig device as an OpenHTF plug — a live, already connected driver instance shared across all phases.
  6. Streams the run live to the dashboard (phases, measurements, prompts, artifacts) as it runs, and persists the final record to your database.

Key consequences you must design around:

  • Device instances are server-owned and long-lived. They are constructed once at rack boot and reused across every phase and every run. Plug tearDown is a deliberate no-op — OpenHTF never disconnects hardware between phases. State you set in phase 1 (an output enabled, a curve programmed) is still there in phase 2, and still there in the next run. This is why teardown discipline (§6) matters and why resume-from-phase (§8) works.
  • The plan is derived from your measurements, not written by hand. The phase list + each phase's @measures limits become the plan shown in the UI/CLI before the run starts. Put display truth in the measurement, never in prose.
  • The trust surface is fixed. Only the names in §3 are injected. Everything else is stock Python.
  • The process is long-lived; only the namespace is fresh. Each run is exec()d in a fresh namespace, but inside the same rack interpreter. Anything that escapes the namespace persists across runs: a background thread you start survives an abort (and still holds its port next run), and imported helper modules stay cached in sys.modules, so edits to a sibling helper need a rack restart to take effect. Stop background work — servers, threads, sockets — in the phase's finally (§6). A test may open a listening port (the "outbound-only" claim in §1 is about the Guppi service, not your OS), but it must close it in finally for the same reason.

1. Environment & versioning#

Thing Value
Python >= 3.12
OpenHTF rack pins openhtf>=1.0; agent pins openhtf>=1.6
Transport pyvisa + pyvisa-py (pure-Python VISA; no NI-VISA needed on a Pi)
Where tests live rig workspace tests/*.py (e.g. /home/guppi/workspace/tests/ on a Pi)
Where hardware is declared rig_config.yml (installed at ~/.guppi/rig_config.yml)

To run what you write: save it as tests/<name>.py in the rig workspace, then run it from the dashboard's test panel or with guppi run <name> (a path to a .py file also works) — see the CLI reference. Results come back via guppi results or the dashboard.

Two runtime modes, transparent to the script:

  • Local ("Pi") mode — agent backend, LAN-only, no auth, no AI.
  • Cloud mode — sign in at guppi.app; the AI agent can drive runs and answer prompts.

The script you write is identical in both. The rack is outbound-only (dials out, never accepts inbound), so you never think about ports or IPs — a rig is addressed by its rig id.


2. The contract (the smallest correct test)#

import openhtf as htf
from openhtf import measures
from openhtf.util import units

@htf.plug(PSU=PSU1)
@measures(htf.Measurement("vout").with_units(units.VOLT).in_range(4.9, 5.1))
def check_rail(test, PSU):
    PSU.set_voltage(1, 5.0)
    PSU.set_output(1, True)
    test.measurements.vout = PSU.measure_voltage(1)

TEST_PHASES = [check_rail]

Three rules, and that's the whole contract:

  1. TEST_PHASES — a module-level flat list of plain phase functions. That list is the test. Nothing else is required — and nothing else is accepted: the plan validator derives the plan from phase names, so htf.PhaseGroup, subtests, and OpenHTF teardown hooks fail validation ("phase names differ") even though stock OpenHTF supports them. This is why abort-safe cleanup is a try/finally inside the phase (§6), not a framework teardown hook.
  2. Bind hardware with @htf.plug(NAME=DEVICE_ID) — the config device id, uppercased, is injected as a plug class. A device with id: psu1 in rig_config.yml is available as PSU1; id: load1 → LOAD1; id: sas1 → SAS1. The left-hand name (PSU=) is just the parameter name inside your function.
  3. Declare limits in @measures — .in_range(lo, hi), .equals(x), .with_units(units.VOLT), etc. The rack introspects these to derive the plan. Building htf.Test(*TEST_PHASES) never executes a phase — but getting TEST_PHASES out of your file means exec'ing the whole module, so the module body runs at plan derivation, off-hardware. Keep top-level code side-effect free: no file I/O, nothing that can raise. A missing data file read at module level fails plan derivation with an opaque load error; deferred into a setup phase it fails there with a clean, named failure.

What a plug actually is#

Inside your phase, PSU is the live, already-connected driver object — you call its Python methods directly (PSU.set_voltage(1, 5.0)), not through any RPC or capability indirection. There is exactly one instance per device, shared across phases and runs.


3. Injected vocabulary — the entire trust surface#

The executor's _build_script_namespace injects exactly this and nothing more:

Name Signature Use
openhtf / htf — the OpenHTF module
time — convenience (time.sleep(...))
PSU1, LOAD1, … — bound device plug classes (uppercased config ids)
capture_artifact (path, waveform, phase=None) -> ref publish a uniform numeric series by reference (a scope trace, or one metric of a sweep), stamped with the run's execution_id. waveform must be {"v": [floats], "sample_rate": float, "t0"?: float, "unit"?: str}; only v is stored (compact f32) and the axis is rebuilt from (t0, sample_rate, n). Any dict lacking "v" raises KeyError. Raises if no artifact store is configured.
safe_shutdown () -> None de-energize every output on the rig. Use in an L2 teardown phase.
arm_guard (path, op, threshold, min_duration=0.0) -> ids arm a run-scoped watchdog abort-limit (L1). The executor disarms exactly what it armed when the run ends — pass, fail, abort, or timeout (§6).
grid (**axes) -> list[point] build a Cartesian setpoint matrix
sweep (points, apply, measure, *, settle_s=0.0, on_point=None) -> rows drive the matrix: command → settle → measure
prompt (text, kind="confirm") -> Any pause the phase and ask a human/agent

__builtins__ is present, so normal imports work (you can from devices.psu.keysight_mp4300 import SasCurve — see §9). Keep that power for data types and stdlib; don't reach around the injected safety vocabulary.

Two more facts about the namespace: __file__ is set to the script's path, so Path(__file__).parent finds data files shipped next to the test; and the script's own directory is put on sys.path, so a sibling helper module (from orbit_parser import ... next to your test) imports directly. Mind the flip side of that convenience: sibling imports are cached in sys.modules for the life of the rack process (§0), so after editing a helper you must restart guppi rack — re-running the test is not enough.


4. Measurements — the pass/fail contract#

Measurements are how a phase declares intent and renders pass/fail. Declare the limit in the decorator; assign the value in the body.

@measures(
    htf.Measurement("vout").with_units(units.VOLT).in_range(4.9, 5.1),
    htf.Measurement("iout").with_units(units.AMP).in_range(0, 0.5),
    htf.Measurement("recovered").equals(True),
)
def characterize(test, PSU):
    test.measurements.vout = PSU.measure_voltage(1)
    test.measurements.iout = PSU.measure_current(1)
    test.measurements.recovered = PSU.measure_voltage(1) > 4.5

Rules that keep results clean:

  • Keep measurements scalar. A measurement value is coerced to a JSON scalar (str() otherwise). Do not assign a list/matrix/waveform to a measurement — it flattens to a string. Reduce to a scalar (min, max, a boolean), and put the full data on the artifact plane (§7).
  • One limit, one place. If the panel should show 4.9–5.1 V, that range lives in .in_range(4.9, 5.1), not in a comment or a prompt string.
  • Name measurements stably. The plan validator compares phase names (ordered) and per-phase measurement-name sets between the derived plan and the running script; a rename mid-approval trips a loud mismatch, not a silent divergence.

5. Prompts — collaborative pause points#

prompt() blocks the phase until answered from the CLI (guppi run asks inline), the web UI (a card on the test panel), or the AI agent. Every prompt — question, answer, responder, timestamps — is recorded in prompts[] of the run-state doc.

prompt("Confirm the DUT is bolted to the cold plate")        # kind="confirm" → True
soc = prompt("Enter the measured pack SoC (%)", kind="input") # kind="input"   → the string
prompt("Trigger the transient, then continue", kind="action") # kind="action"  → True

Guidance:

  • Use prompts only where the test genuinely needs a human — bench setup, a manual stimulus, a value only the operator can read. An unattended run should fly through a prompt-free script.
  • guppi run --step inserts a confirm-gate before every phase as a debug overlay; those gates never appear in the plan or the result phases.
  • A prompt raises RuntimeError if the run is aborted while waiting, so it cooperates with abort.

6. Safety — the three layers (design for all three)#

Never rely on software to catch a microsecond fault. Layer your protection:

Layer Mechanism Where it lives
L0 the instrument's own current/OCP/OVP limit set it first in your setup phase, or in rig_config.yml device kwargs
L1 watchdog abort-limit — samples telemetry, de-energizes + latches on breach rig_config.yml safety.abort_limits, or per-run arm_guard(...)
L2 explicit de-energize on the way out safe_shutdown() in a try/finally inside every phase that energizes

Canonical shape of a phase that energizes:

@htf.plug(PSU=PSU1)
@measures(htf.Measurement("peak_current"))
def stress(test, PSU):
    try:
        PSU.set_current(1, 5.0)                       # L0: hardware current limit first
        arm_guard("psu1.1.current", ">", 5.5, min_duration=0.2)  # L1: run-scoped backstop
        PSU.set_voltage(1, 5.0)
        PSU.set_output(1, True)
        time.sleep(0.5)
        test.measurements.peak_current = PSU.measure_current(1)
    finally:
        safe_shutdown()                              # L2: de-energize no matter what

Why the finally must live inside the phase — abort semantics. A manual abort raises an exception into the currently running phase's thread: that phase's finally blocks execute, but every later phase in TEST_PHASES is skipped — including a trailing teardown phase. The same applies to a phase timeout (§8). So a standalone teardown phase is the normal-completion off-path only; the abort off-path is the try/finally inside the energizing phase itself. Keep that finally short and per-op fault-tolerant (wrap each de-energize call in its own try/except): if the run doesn't finish within ~10 s of an abort, the executor escalates by re-killing the phase thread every few seconds — a finally wedged in slow driver I/O gets interrupted too. Two related facts: a guard trip (L1) de-energizes the rig before aborting the run, so it doesn't depend on your finally; and a hard kill (SIGKILL, power loss) bypasses Python entirely — which is why L0 hardware limits are non-negotiable.

Notes:

  • arm_guard is run-scoped and executor-disarmed. It records the ids it armed, and the executor disarms exactly those when the run ends — in a finally, so it happens on pass, fail, abort, or timeout alike. A guard never leaks into the next run and you never write disarm code. Persistent guards belong in rig_config.yml.
  • Thresholds are signed for bidirectional supplies. A regenerative supply sinking current reads negative — arm one limit per direction (e.g. >, 80 for source OC and <=, -40 for sink OC).
  • safe_shutdown() invokes every device's energizing=True capability with {"enabled": False} — a guaranteed rig-wide de-energize.

7. Performance — measure fast signals from the instrument's buffer#

A polled read is capped by the SCPI round-trip (1–50 ms, serialized per instrument) plus the instrument's NPLC integration (~16–20 ms). Realistically a few Hz to ~100 Hz for a whole device. Polling faster does not sample faster — it returns duplicates.

For anything fast (load transients, inrush, ripple), do not poll. Arm the instrument's hardware-timed buffer, let it capture on its own clock, then read the trace back in one transfer and publish it by reference:

@htf.plug(SCOPE=SCOPE1)
def capture_transient(test, SCOPE):
    wave = SCOPE.fetch_waveform(channel=1)      # hardware did the DAQ
    capture_artifact("scope1.transient", wave, phase="capture_transient")

The waveform rides the artifact plane (compact f32 + summary stats), joined to the run by execution_id; it never bloats the result JSON. Reserve polled bench reads for the slow, precise tier (setpoints, NPLC-accurate DMM values).

Budget per-point SCPI cost in a paced sweep. Driver methods are not equal. On the MP4300: a V or I read is one query each; measure_power(ch) is measure_voltage() × measure_current() (the mainframe has no power query), so asking for v, i, and p costs four queries with V and I each read twice — and not coherently. Read V and I once and compute p = v * i in the script (§9.6 does this). A set_current_scale is one write; a set_sas_curve reprogram is the most expensive per-point op — a four-command compound write plus SYST:ERR? drains before and after. Keep the per-step wall budget comfortably above the summed per-step SCPI cost (on a Pi driving two SAS channels with a curve reprogram per step, ~1 s steps overrun; ~5 s holds pace), and skip idempotent writes when the setpoint hasn't meaningfully moved (a small deadband on the swept variable).

The capture_artifact contract (know this exactly). The helper accepts a single uniform numeric series, shaped like what fetch_waveform() returns:

{"v": [ ... floats ... ],   # required: the values (the only thing stored)
 "sample_rate": <float>,    # required: samples per axis-unit
 "t0": <float>,             # optional: axis start  (default 0.0)
 "unit": "<str>"}           # optional: engineering unit of v

Only v is encoded (compact f32); the axis is reconstructed from (t0, sample_rate, n), so the stored series is always 1-D and uniformly spaced. v must be a list of floats. A dict without "v" raises KeyError: 'v' — you cannot hand capture_artifact a {"rows": ...} matrix, an arbitrary JSON blob, or a list of dicts. To persist a multi-metric sweep, publish one series per metric (§8). Scalars and metadata belong in measurements, not the artifact.


8. Sweeps and resume#

Declarative sweeps#

Use grid + sweep instead of hand-writing nested loops — you get a structured, dimensioned result and the loop can't drift from the plan.

def series(rows, key):
    """One metric column from sweep() rows, as a plain float list."""
    return [float(r["values"][key]) for r in rows]


@htf.plug(PSU=PSU1, LOAD=LOAD1)
@measures(htf.Measurement("vout_min").with_units(units.VOLT).in_range(4.9, 5.1))
def characterize(test, PSU, LOAD):
    rows = sweep(
        grid(vin=[20, 28, 36], iout=[0.5, 1.0]),   # row-major, first axis outer
        apply=lambda p: (PSU.set_voltage(1, p["vin"]), LOAD.set_current(1, p["iout"])),
        measure=lambda p: {"vout": PSU.measure_voltage(1)},
        settle_s=0.2,                               # wait between apply and measure
    )
    # rows = [{point, values, t_ms}, ...]. Persist ONE numeric series per metric —
    # capture_artifact stores a uniform 1-D array, not a matrix (§7).
    capture_artifact(
        "sweep.characterize.vout",
        {"v": series(rows, "vout"), "sample_rate": 1.0, "t0": 0.0, "unit": "V"},
        phase="characterize",
    )
    test.measurements.vout_min = min(r["values"]["vout"] for r in rows)

sweep is sequential by construction (command → settle → measure) — the correct, reproducible ordering for qualification. An optional on_point= callback is invoked after each point with (point, values) — useful for a progress log or an early-exit check between steps. Do not assign the row list to a measurement (§4), and do not hand it to capture_artifact as {"rows": …} — that raises KeyError: 'v' (§7). Persist each metric as its own series.

Choosing the series axis. capture_artifact reconstructs a uniform axis from (t0, sample_rate, n):

  • 1-D sweep over a uniform axis — encode the real axis. For grid(irradiance_pct=[20, 40, 60, 80, 100]) the step is 20, so t0=20.0, sample_rate=1.0/20.0 (samples per %-irradiance) rebuilds 20, 40, 60, 80, 100.
  • Multi-axis grid (or a non-uniform axis) — there is no single uniform axis, so publish against the row index: t0=0.0, sample_rate=1.0. Rows stay in grid's row-major order (first axis outer), and the swept setpoints are already captured per row in rows[i]["point"].

Sweep points stream to telemetry as they happen#

sweep doesn't just return rows — as each point completes it publishes every numeric key of the measure() return dict, and every key of the point dict, as live telemetry channels (under a per-phase path). Non-numeric values are skipped (bools become 0/1). Telemetry is persisted continuously with no retention cap, so an abort at hour N leaves every point up to that instant in the record — which makes telemetry, not a capture_artifact at the end of the phase, the right persistence for a long or abortable sweep (the artifact only exists if the phase reaches the capture_artifact call). Three consequences: keep helper/bookkeeping state out of point dicts (every key becomes a channel); name commanded and measured values differently (vout_set vs vout) — point and values keys share one channel namespace, and a collision silently interleaves setpoint and readback on a single channel; and use capture_artifact for what it's for — compact, uniform series you want joined to the run as artifacts, from phases that complete.

Long phases: the default timeout will kill your sweep#

OpenHTF applies a default phase timeout of 180 s. A phase that runs longer — any real sweep or replay — is terminated with outcome TIMEOUT, and a timed-out phase skips all remaining phases exactly like an abort (§6), so the in-phase try/finally is the cleanup path here too. Size an explicit timeout above the worst-case wall clock:

@htf.PhaseOptions(timeout_s=N_POINTS * PER_POINT_BUDGET_S + 900)
@htf.plug(SAS=SAS1)
def orbit_replay(test, SAS):
    ...

Robustness on multi-hour runs#

Thousands of SCPI round-trips will include a few transient blips. The codec does not retry for you: a transport timeout raises (pyvisa's VisaIOError; the transport resyncs the link, and after 4 consecutive timeouts drops and reopens the session), while a deterministic rejection — an invalid curve, an out-of-range setpoint — raises ValueError. Pattern that holds up: retry each SCPI call once after a short delay, never retrying ValueError (it will fail identically); on a read that fails even the retry, record a NaN gap and keep sweeping; abort only on ~5 consecutive failures (that's a real comms loss — raise, and let the phase finally de-energize). Your arm_guard watchdogs sample telemetry independently of the script, so L1 protection stays live while the script is retrying.

Resuming a failed run#

A finished run can be re-run from a phase onward — the "resume from here" button on a failed phase, or resume_from_phase on the run command. The rack slices TEST_PHASES from the named phase; because device plugs are server-owned and keep their state across runs, earlier phases needn't repeat. It is a new execution with its own id; the plan still describes the whole test, so skipped phases read as skipped. Design phases so a later one can start from the state a former one left (or make each phase re-establish what it needs).


9. The MP4300 Solar Array Simulator as an irradiance source#

This is the section you came for. The built-in KeysightMP4300 driver turns a Keysight MP4300 mainframe into a photovoltaic source your DUT (an MPPT charger, a power board, a bus converter) sees as a real solar array.

9.1 How irradiance maps onto the SAS#

A PV panel's I–V curve is defined by four points, and irradiance/temperature move them in physically specific ways:

PV parameter Driver field Moves with…
Isc — short-circuit current isc Irradiance — linearly. 1000 W/m² → full Isc; 500 W/m² → ~half Isc. This is your primary irradiance knob.
Imp — current at max-power point imp Irradiance, ~linearly (tracks Isc).
Voc — open-circuit voltage voc Irradiance weakly (logarithmic); temperature strongly (negative tempco).
Vmp — voltage at max-power point vmp Mostly temperature.

So there are two clean ways to drive irradiance, and you'll typically use both:

  1. Current scale factor (fast irradiance knob). set_current_scale(ch, pct) scales the whole programmed curve's current by 1–100 %. Program the curve once at full sun (100 %), then treat the scale percentage as irradiance in percent of STC: 100 % = 1000 W/m², 20 % = 200 W/m². A single SCPI write with no re-validation — ideal for an irradiance sweep or a moving-cloud profile. Pair it with set_voltage_scale(ch, pct) to model the temperature axis (Voc/Vmp) independently.

    The scale factor has a floor. Scaling current by X % also scales the curve's dI/dV slope by X %, and the mainframe enforces a minimum slope of 0.01 A/V (§9.2) — so a valid full-sun curve becomes invalid below a computable scale, and the mainframe silently keeps the old curve. Compute the floor up front: min_scale_pct = 100 × 0.01 / ((isc − imp) / (voc − vmp)). Below it, either reprogram the full curve for that operating point (option 2) or turn the output off — for typical panel curves the floor lands in the low-tens of percent, well inside a dawn/dusk or eclipse profile.

  2. Reprogram the full curve. set_sas_curve(ch, SasCurve(voc, isc, vmp, imp)) writes a new four-point curve atomically. Use this when you want physically correct Voc and Isc for a specific (irradiance, temperature) operating point — e.g. converting a datasheet's per-irradiance curves directly.

9.2 Hard limits — the module capability envelopes#

An MP4300 is a mainframe holding heterogeneous SAS modules; each installed module is one channel, addressed inline as (@ch). At connect the driver reads *RDT? to learn which slots are populated and with which model, then selects that slot's envelope. Curves and limits are validated against the per-slot envelope — never a single mainframe-wide number.

MODULE_LIMITS table (from the MP4300 Operating & Service Guide "Power Module Output Quadrants" + the MP4300A data sheet):

Model Voc/Vmp max Isc/Imp max Power max Quadrants Sink max Ranging
MP4361A 160 V 10 A 1000 W 2-quadrant −10 A autoranging
MP4362A 130 V 8 A 1000 W single (source-only) −0.5 A fixed
MP4351A 160 V 10 A 1400 W 2-quadrant −10 A autoranging
MP4352A 80 V 20 A 1400 W 2-quadrant* −20 A autoranging

* The guide prints "??" for the MP4352A's 2-quadrant cell; the driver treats it as 2-quadrant but confirm sink behaviour on real hardware before relying on it. The default envelope (no detection, no config override) is the MP4361A.

Curve validity rules enforced client-side by validate_sas_curve before the write (a curve that violates these raises ValueError, so you catch it in Python rather than having the mainframe silently reject it):

  • All four parameters >= 0.
  • Vmp <= Voc and Imp <= Isc.
  • Isc >= max(0.1 % of the module's Isc_max, 10 mA) — below the module's minimum representable short-circuit current the mainframe rejects the curve, lights ERR, and silently keeps the previous curve. This is the footgun the client-side check exists to catch.
  • Each of Voc, Vmp, Isc, Imp within the module max.
  • Pmp = Vmp × Imp <= module p_max.

Two more rules the mainframe enforces that validate_sas_curve does NOT check (found on hardware — they surface only through the silent-rejection guard below):

  • dI/dV slope bounds. The slope between (Vmp, Imp) and (Voc, 0) — dI/dV = (Isc − Imp) / (Voc − Vmp) — must satisfy 0.01 A/V <= dI/dV <= 4.153892 A/V. Too shallow is below the module's minimum representable step; too steep exceeds its slew capability.
  • Knee geometry. (Vmp, Imp) must lie above the straight line from (0, Isc) to (Voc, 0): Imp > Isc × (1 − Vmp/Voc). A knee on or below that line is rejected.

A curve that violates either is rejected with ERR while the old curve stays active — exactly the failure mode set_sas_curve's error-drain raises on. Validate these two in your script before writing (they're two lines of arithmetic) rather than discovering them mid-run.

Silent-rejection guard. Even after the client check, set_sas_curve drains SYSTem:ERRor? before and after the write and raises if the mainframe logged a rejection — because an invalid curve is otherwise accepted-looking while the old curve stays active. Trust the exception; don't assume a write "took".

Scale factors: set_current_scale / set_voltage_scale accept 1–100 % only (values outside raise). You cannot scale to 0 % — to kill the array, turn the output off (set_output(ch, False)), don't scale to zero.

Other envelope-bounded setters (each validates against the per-slot module):

  • set_current_limit(ch, A) — positive limit, <= isc_max (voltage priority).
  • set_current_limit_negative(ch, A) — sink limit magnitude, <= sink_current_max. A source-only module (MP4362A) rejects anything beyond −0.5 A.
  • set_voltage_limit(ch, V) / set_voltage_limit_low(ch, V) — <= voc_max.
  • set_power_limit(ch, W) — <= p_max. Caps one module so the summed frame power can't trip the mainframe FPL (which drops all outputs).
  • set_ovp(ch, V), set_ocp(ch, A) — programmable protection. OCP applies in SAS mode; in Fixed mode the mainframe ignores it and enforces the current limits instead.

9.3 Modes, settling, and faults you must handle#

  • Mode: set_mode(ch, "SAS") for curve operation (what you want for irradiance), set_mode(ch, "FIX") for a plain fixed-V/I supply.
  • Regulation priority (Fixed mode only): set_priority_mode(ch, "VOLT"|"CURR").
  • SAS compensation bandwidth: set_sas_bandwidth(ch, mode) where mode is DEFAULT | FAST_LOWC | FAST_HIGHC | SHUNTSW. Match this to your DUT converter's input capacitance/topology — the wrong mode can oscillate the SAS loop. For an MPPT DC-DC charger, FAST_LOWC/FAST_HIGHC are the usual choices; DEFAULT is the widest-tolerance fallback.
  • Settling: command ack is ≤1 ms but the output settles in ~100 ms. Before a dependent readback, call PSU.wait_operation_complete() (*OPC?) or time.sleep(0.1). The telemetry sampler polls continuously and doesn't need it.
  • Realised vs requested curve: the module fits an exponential model, so the realised Vmp/Imp can differ from what you asked (Isc/Voc are exact). Read back what the DUT actually sees with get_actual_sas_curve(ch).
  • Faults / latches:
    • channel_conditions(ch) decodes live status to named flags (OV, OC, OT, CV, CC, FLT, …).
    • clear_output_protection(ch) clears most trips (OVP/OCP/OT) — but FLT is a hardware-fault latch that only a power cycle clears. The driver logs if FLT is still set after a clear.
    • Frame-level faults (frame_faulted telemetry signal / frame_conditions()): fan, over-temp, missing lockout bar, input power fail, or FPL (summed power exceeded) — any of these takes down every channel.
    • Blocking diode: set_diode_mode(ch, True) inserts a reverse-current blocking diode; with it engaged the channel cannot sink current, negative limits/setpoints are invalid, and voltage readback carries a ~0.7 V anode offset. Switching it reverts the module to *RST (drops the output) — set it in setup, not mid-run.

9.4 Telemetry signals the MP4300 streams#

Per installed channel: {ch}.voltage, {ch}.current, {ch}.output, {ch}.voltage_setpoint, {ch}.current_setpoint, {ch}.protection_tripped, {ch}.diode_mode, {ch}.thermal_margin_c. Rig-wide: frame_faulted.

thermal_margin_c is worth watching in a long irradiance-at-full-power run: the module derates current 1 %/°C above 40 °C ambient, so a shrinking margin predicts an OT trip before it happens — arm an arm_guard on it and back off rather than react to the trip. Abort-limit paths use the sasN.<signal> form (e.g. sas1.1.current, sas1.frame_faulted).

9.5 rig_config.yml for the SAS#

devices:
  - id: sas1                      # → plug SAS1 in tests
    name: "MP4300 Solar Array Sim"
    type: KeysightMP4300
    enabled: true
    num_channels: 2               # declare installed count; *RDT? confirms at connect
    rate_hz: 5                     # SAS health moves slowly; a few Hz is plenty
    connection:
      type: VISA
      address: "TCPIP0::192.168.1.50::inst0::INSTR"
      timeout: 10.0
    # Optional: pin the envelope explicitly to DISABLE auto-detection.
    # Omit to let *RDT? pick each slot's module envelope automatically.
    # module_limits: { voc_max: 160, isc_max: 10, p_max: 1000 }

safety:
  abort_limits:
    - { path: sas1.1.current, op: ">=", threshold: 10, min_duration: 0.1,
        label: "sas1 ch1 over-current" }
    - { path: sas1.frame_faulted, op: ">=", threshold: 1,
        label: "sas1 mainframe fault" }

If you pin module_limits or channel_limits in config, that is authoritative and detection leaves the bounds alone (mirrors the RP5900 driver). Omit them to let *RDT? select the correct per-slot envelope.

9.6 End-to-end: an irradiance sweep on a DUT#

This programs a full-sun curve once, then steps irradiance via the current scale factor and records how the DUT's harvested power tracks — the canonical MPPT / power-board qualification.

"""Irradiance sweep — hold a fixed panel curve, step irradiance 20→100 %,
and verify the DUT keeps harvesting near the max-power point at each level."""

import openhtf as htf
from openhtf import measures
from openhtf.util import units
from devices.psu.keysight_mp4300 import SasCurve   # __builtins__ import allowed

CH = 1
# One panel's STC (full-sun) curve. Within the MP4361A envelope
# (160 V / 10 A / 1 kW): Pmp = 32.4 V * 8.6 A = 279 W. OK.
STC = SasCurve(voc=38.0, isc=9.1, vmp=32.4, imp=8.6)

# Irradiance axis: 20 % steps starting at 20 %. Encoding it as (t0, sample_rate)
# lets each published series carry the real x-axis (see §8).
IRR_STEP_PCT = 20.0
IRR_AXIS = {"t0": IRR_STEP_PCT, "sample_rate": 1.0 / IRR_STEP_PCT}  # samples per %


def series(rows, key):
    """One metric column from sweep() rows, as a plain float list."""
    return [float(r["values"][key]) for r in rows]


@htf.plug(SAS=SAS1)
def setup_array(test, SAS):
    # Order matters: mode → compensation → curve → protection → enable.
    SAS.set_mode(CH, "SAS")
    SAS.set_sas_bandwidth(CH, "FAST_LOWC")     # match the DUT's MPPT input stage
    SAS.set_sas_curve(CH, STC)                 # raises if rejected — trust it
    SAS.set_current_scale(CH, 100.0)           # start at full sun
    SAS.set_ocp(CH, 9.5)                        # L0 above Isc, below module max
    SAS.set_power_limit(CH, 350.0)             # keep the frame under FPL
    arm_guard("sas1.frame_faulted", ">=", 1)   # L1 backstop on frame health
    SAS.set_output(CH, True)
    SAS.wait_operation_complete()


@htf.plug(SAS=SAS1)
@measures(
    # Reduce the matrix to scalar pass/fail; the full matrix goes to an artifact.
    htf.Measurement("harvest_at_full_sun_w").with_units(units.WATT).in_range(240, 300),
    htf.Measurement("min_harvest_ratio").in_range(0.85, 1.0),
)
def irradiance_sweep(test, SAS):
    def measure(p):
        # Read V and I once each and derive P — measure_power() is V×I under
        # the hood, so asking for all three costs 4 queries, V and I twice.
        v = SAS.measure_voltage(CH)
        i = SAS.measure_current(CH)
        return {
            "v": v, "i": i, "p": v * i,
            # Available max power at this irradiance = scaled Pmp of the curve.
            "p_avail": STC.pmp() * p["irradiance_pct"] / 100.0,
        }

    try:
        rows = sweep(
            grid(irradiance_pct=[20, 40, 60, 80, 100]),
            apply=lambda p: SAS.set_current_scale(CH, p["irradiance_pct"]),
            measure=measure,
            settle_s=0.3,   # SAS output settle + DUT MPPT re-track
        )
    finally:
        # L2 *inside the energizing phase*: a manual abort raises into the
        # running phase and SKIPS every later phase — including the teardown
        # below — so this finally is the abort off-path.
        safe_shutdown()
    # One uniform series per metric (§7); the irradiance axis rides (t0, sample_rate).
    for metric, unit in (("p", "W"), ("v", "V"), ("i", "A"), ("p_avail", "W")):
        capture_artifact(
            f"sas1.irradiance_sweep.{metric}",
            {"v": series(rows, metric), "unit": unit, **IRR_AXIS},
            phase="irradiance_sweep",
        )

    full_sun = next(r for r in rows if r["point"]["irradiance_pct"] == 100)
    test.measurements.harvest_at_full_sun_w = full_sun["values"]["p"]
    test.measurements.min_harvest_ratio = min(
        r["values"]["p"] / r["values"]["p_avail"] for r in rows
    )


@htf.plug(SAS=SAS1)
def teardown(test, SAS):
    # Normal-completion off-path ONLY: an abort (or phase timeout) skips this
    # phase entirely — the sweep's finally above is the abort off-path (§6).
    safe_shutdown()

TEST_PHASES = [setup_array, irradiance_sweep, teardown]

Why it's built this way:

  • Curve programmed once, irradiance via scale — one SCPI write per step, physically correct (Isc scales with irradiance). Valid here because the lowest step (20 %) stays above this curve's scale floor (100 × 0.01 / ((9.1 − 8.6)/(38.0 − 32.4)) ≈ 11 % — §9.1); sweeping lower would need per-point curve reprogramming or output-off below the floor.
  • safe_shutdown() in the sweep phase's finally — the abort off-path; the trailing teardown phase only runs on normal completion (§6).
  • FAST_LOWC compensation chosen for an MPPT DC-DC input — prevents SAS-loop oscillation. Change to match your actual converter.
  • settle_s=0.3 covers both the SAS output settle (~100 ms) and the DUT's MPPT re-tracking after each irradiance step — tune to your DUT.
  • Scalar measurements, one series per metric — the sweep rows would flatten to a string if assigned to a measurement, so each metric (p, v, i, p_avail) rides the artifact plane as its own uniform series, and only the reduced scalars (harvest_at_full_sun_w, min_harvest_ratio) carry limits. capture_artifact stores a 1-D array, never a {"rows": …} matrix (§7).
  • All three safety layers: OCP/power-limit (L0), arm_guard on frame health (L1), safe_shutdown() in the energizing phase's finally (L2).

9.7 Modeling irradiance and temperature together#

If you need a full (irradiance, temperature) operating grid with physically correct Voc and Vmp at each point, reprogram the curve per point instead of scaling. Precompute the four points from your panel model / datasheet:

@htf.plug(SAS=SAS1)
@measures(htf.Measurement("min_harvest_ratio").in_range(0.85, 1.0))
def irradiance_temp_grid(test, SAS):
    rows = sweep(
        grid(irradiance_wm2=[400, 700, 1000], cell_temp_c=[15, 45, 70]),
        apply=lambda p: SAS.set_sas_curve(CH, panel_curve(p["irradiance_wm2"],
                                                           p["cell_temp_c"])),
        measure=lambda p: {"p": SAS.measure_power(CH),
                           "p_avail": panel_curve(p["irradiance_wm2"],
                                                  p["cell_temp_c"]).pmp()},
        settle_s=0.4,
    )
    # A 2-D grid has no single uniform axis, so publish each metric against the
    # row index (t0=0, sample_rate=1). Rows stay row-major — irradiance outer,
    # temperature inner — and each row's setpoints live in rows[i]["point"].
    # `series` is the helper from §8/§9.6.
    idx_axis = {"t0": 0.0, "sample_rate": 1.0}
    capture_artifact("sas1.irr_temp_grid.p",
                     {"v": series(rows, "p"), "unit": "W", **idx_axis},
                     phase="irradiance_temp_grid")
    capture_artifact("sas1.irr_temp_grid.p_avail",
                     {"v": series(rows, "p_avail"), "unit": "W", **idx_axis},
                     phase="irradiance_temp_grid")
    test.measurements.min_harvest_ratio = min(
        r["values"]["p"] / r["values"]["p_avail"] for r in rows)

where panel_curve(irradiance, temp) returns a validated SasCurve (Isc ∝ irradiance; Voc down with temperature via the panel tempco). Each write is atomically validated against the module envelope, so an out-of-envelope (irradiance, temp) point fails loudly at that step.

9.8 Pace a timed slew against a deadline, not a fixed sleep#

A slew that re-programs the curve once per second over a fixed duration must pace against an absolute deadline. Every device write is a blocking round-trip to the instrument, so a fixed sleep per step adds to that latency rather than absorbing it:

# WRONG — wall-clock = DURATION_S + Σ(write latency). With 2 channels
# (2 writes/step) this can run a 300 s sweep in ~10 min on real hardware.
for k in range(steps):
    apply_curve(k)
    time.sleep(STEP_S)

# RIGHT — latency is absorbed; total ≈ DURATION_S.
t0 = time.monotonic()
for k in range(steps):
    apply_curve(k)
    time.sleep(max(0.0, t0 + (k + 1) * STEP_S - time.monotonic()))

If a step's writes exceed STEP_S the sleep goes non-positive and the loop proceeds — the true floor is steps × write-latency, so raise STEP_S if that dominates. This changes pacing only: the curve values per step are unchanged, so the physical sweep is identical, just on schedule.


10. Bench pitfalls — assume nothing, verify everything#

The failures that cost real debugging time are the silent ones: every command is accepted, nothing errors, and the result is still wrong. They share a root cause — assuming instead of verifying — and a cure: make each link in the chain prove itself in the results.

  • Power-up order is part of the test. Energize source → DUT → sink, and tear down in reverse. A DUT with its own power management often refuses to arm without its supply — silently. Because teardown de-energizes the bench between phases, every phase must rebuild this chain, not just the first.
  • Program the source fully. Set voltage and current limit in the script; a limit inherited from the front panel folds the supply back mid-sweep with no error.
  • Read back actuations. Commanding an output on is not the same as it engaging. Record a readback, and confirm the stimulus physically happened before judging the DUT against it.
  • Interrogate instrument config before energizing. Panel-set state — a load's turn-on voltage (Von), slew rates, protection setpoints, inherited limits — can silently veto a test with no error anywhere. It is all readable over the wire: set what you depend on explicitly, and treat anything you didn't set as unknown, not default.
  • Prove state by measurement, not status flags. Device telemetry can lag or mis-report; record it as a diagnostic and let physics — does current actually flow? — carry the verdict. Poll for state changes with a deadline, never a fixed sleep.
  • When a step does nothing and nothing errors, look upstream. Supply, limit, mode, readbacks — in that order — before suspecting the DUT.

11. What "up to Guppi standards" means#

A finished test has TEST_PHASES as a module-level list of phase functions, every device bound with @htf.plug(NAME=DEVICE_ID) using ids that match rig_config.yml, and every pass/fail limit in an @measures(...) decorator rather than prose. Measurements stay scalar; each waveform or sweep metric goes through capture_artifact as its own uniform series ({"v", "sample_rate", …}), never a {"rows": …} matrix.

Safety is layered, in order: an energizing phase sets L0 hardware limits first, arms an L1 arm_guard, and de-energizes in an L2 try/finally inside the phase itself — an abort or phase timeout skips later phases, so a separate teardown phase is the normal-completion path only (§6). Sources always have voltage and current limits programmed, power-up order is explicit in every phase (source → DUT → sink, reversed on teardown), and actuations are proven by measurement — does current actually flow? — not by DUT status flags (§10).

Structurally: fast signals come from a hardware buffer, never a polling loop; test matrices use grid/sweep, not hand-rolled nested loops; prompts appear only where a human is genuinely required; and phases are resume-safe — a later phase either works from the state a former one left or re-establishes what it needs.

For the MP4300 specifically: configure mode → bandwidth → curve → protection → enable, in that order; keep curves inside the per-slot module envelope; trust set_sas_curve exceptions; call wait_operation_complete() before dependent readbacks; and watch frame_faulted / thermal_margin_c on long runs.